Table 3:
Acceptability and Reproducibility Criteria: Summary
Acceptability Criteria
Reproducibility Criteria
Individual spirograms are "acceptable" if:
  They are free from artifacts
 
Cough or glottis closure during the first second of exhalation
Eary termination or cutoff
Variable effort
Leak
Obstructed mouthpiece
  Have good starts
 
Extrapolated volume less than 5% of FVC or 0.15 L, whichever is greater; OR
Time-to-PEF of less than 120 ms (optional until further information is available)
  Have a satisfactory exhalation
 
6 s of exhalation and/or a plateau in the volume-time curve; OR
Reasonable duration or a plateau in the volume-time curve; OR
If the subject cannot or should not continue to exhale
After 3 acceptable spirograms have been obtained, apply the following tests:
Are the two largest FVC within 0.2 L of each other?
Are the two largest FEV1 within 0.2 L of each other?
If both of these criteria are met, the test session may be concluded.
If both of these criteria are not met, continue testing until:
Both of the criteria are met with analysis of additional acceptable spirograms; OR
A total of eight tests have been performed; OR
The patient/subject cannot or should not continue
Save at a minimum the three best manuvers
Adapted from reference 4
Copyright 2003 The Cleveland Clinic Foundation

Return to Pulmonary Function Testing Chapter