| Table 4: | ||
|
Equipment
Quality Control Summary
|
||
|
Test
|
Minimum
Interval
|
Action
|
| Volume |
Daily
|
3-L syringe check |
| Leak |
Daily
|
3 cm H20 constant pressure for 1 min |
| Linearity |
Quarterly
Weekly (flow spirometers) |
1-L increments with a calibrating syringe measured over entire volume range (flow spirometers simulate several different flow ranges) |
| Time |
Quarterly
|
Mechanical recorder check with stopwatch |
| Software |
New
versions
|
Log installation date and perform test using "known" subject |
| Adapted from reference 4. | ||
|