Table 4:
Equipment Quality Control Summary
Test
Minimum Interval
Action
Volume
Daily
3-L syringe check
Leak
Daily
3 cm H20 constant pressure for 1 min
Linearity
Quarterly
Weekly (flow spirometers)
1-L increments with a calibrating syringe measured over entire volume range (flow spirometers simulate several different flow ranges)
Time
Quarterly
Mechanical recorder check with stopwatch
Software
New versions
Log installation date and perform test using "known" subject
Adapted from reference 4.